47A High-Resolution Spectroscopy

Scientific Scope

TPS 47A High-Resolution Spectroscopy Beamline is a high-brightness hard X-ray spectroscopy beamline constructed as one of the Phase III beamline at the Taiwan Photon Source (TPS) of the National Synchrotron Radiation Research Center (NSRRC). The beamline is specifically designed for advanced high-resolution synchrotron spectroscopy and next-generation materials research. TPS 47A beamline design enables simultaneously high photon flux, high energy resolution, and micron-scale beam focusing capability, meeting the demanding requirements of next-generation high-resolution hard X-ray spectroscopy experiments.

TPS 47A consists of two core endstations: the A1 photo-in/photo-out Spectroscopy station (HERFD-XAS,XES,RIXS), and the A2 Hard X-ray Photoelectron Spectroscopy station (HAXPES). Both endstations are dedicated to high-resolution spectroscopy and advanced materials applications. High-resolution spectroscopy plays a critical role in modern materials science, semiconductor technology, and quantum-material research. It has become an indispensable tool for next-generation semiconductor, energy, and advanced materials research.

Beamline Description

The beamline employs an undulator insertion device to generate a high-brightness photon source, combined with a Double Crystal/Double Multilayers Monochromator (DCM/DMM), a 4-bounce High-Resolution Monochromator (HRM), and Kirkpatrick–Baez (KB) focusing mirrors. This configuration provides high photon flux, high energy resolution, and micron-scale beam focusing capability simultaneously, meeting the requirements of next-generation high-resolution hard X-ray spectroscopy experiments.

Specifications

Source:

ID, 3 m IUT24

Monochromator:

Crystal monochromator

two sets of crystals, Si (111) and Si (311), switchable

Multilayer monochromator

Energy Range:

3.2 - 20 keV

Resolution or Resolving Power:

1.86 × 10-4 (Si 111) / 9.71 × 10-5 (Si 311) by DCM

High-Resolution Monochromator

4-bounce channel cut crystal monochromator

Si (111), Si (110) and Si (100)

Resolution or Resolving Power:

50 meV (Si 444), 200 meV (Si 400), 320 meV (Si 220) @ 8 keV

A1 HERFD XAS/ XES endstation

Focused Spot Size:

20 x 20 μm2 

Flux Range (ph/s):

2 x 1014 photons/sec at 6 keV

Detector

EIGER 2 S 1M-L, SDD, PIN diode

analyzer

1 m radius multi-analyzer Johann-type setup

A2 HAXPES endstation

Focused Spot Size:

6 x 6 μm2 

Flux Range (ph/s):

7.2 x 1013 photons/sec at 6 keV

Detector

2D CMOS with true pulse counting

analyzer

SPECS PHOIBOS 150 HV NAP

IUT24 Energy range

Bandwidth of DCM

Find a Beamline

47A Home

Optical Layout

Endstations

Gallery



Techniques

  • 0.7–2 Å MAD/SAD phasing

Disciplines

  • X-ray optics

General Information

  • Source: IUT24
  • Energy Range: 3.2–20 keV
  • Focused Spot Size: 20 x 20 μm²
  • Status: under construction